User-friendly compact design
Based on over 40-year experience in manufacturing hematology analyzers, Nihon Kohden’s Celltac α provides stress-free operation with newly designed hardware and software interface.
Flexible aspiration modes
Celltac α has several blood aspiration modes. The cap pierce mode (available on MEK-6500J/K Series) contributes to safer operation by eliminating the chance of touching blood.
Large data capacity
Celltac α uses SD cards for reliable data storage. Over 15,000 results can be stored.
Hemathology Analyzer
Model : MEK- 6510K
Manufacturer : Nihon Kohden
Origin : Japan
Specifications
Parameters WBC, LY%, MO%, GR%, LY, MO, GR, RBC, HGB, HCT, MCV, MCH, MCHC,
RDW-CV, RDW-SD, PLT, PCT, MPV, PDW
Measurement methods WBC, RBC, PLT – impedance method
+ HCT – calculation from histogram
+ HGB – cyanmethemoglobin optical detection
+ WBC differential – calculation from histogram
Measurement modes Open
+ Closed
+ Pre-dilution
+ WBC high
+ WBC low
Counting time 60 sec/sample (Open mode)
Sample volume aspirated : 30 μL (CBC) for normal measurement mode; 10 or 20 μL for pre-dilution mode
Data storage Over 15,000 results including histograms can be stored on an SD card.
LCD display 5.7 inch 240 × 320 dots TFT-LCD
Dimensions 230 W × 450 D × 428 H (mm)
Weight Approx. 20 kg