MEDUCORE Standard supports you every step of the way – the defibrillation and monitoring unit guides you intuitively through resuscitation. Clear voice prompts, a metronome for maintaining the correct chest compression rate and a professional alarm system ensure you always have a clear overview.
Simultaneous shock charging while automatically analyzing the cardiac rhythm in AED mode means that, if required, shock administration can take place immediately so as to ensure chest compression is interrupted by defibrillation for as short a period as possible.
The large and bright color display means you have the key patient parameters in sight at all times – as well as, depending on your requirements, SpO2 measurement, manual defibrillation mode and 6-lead ECG.
WEINMANN Emergency as a matter of course
MEDUCORE Standard in combination with a MEDUMAT ventilator offers an innovative world first: configured to meet your needs and mounted on the LIFE-BASE portable system, the two devices create an individual unit consisting of defibrillation, monitoring and ventilation.
Your benefits at a glance:
Clear overview – large color display shows key vital parameters
Intuitive – simple to operate and easy menu guidance
Safe – clear alarm system
Strong – rugged design
Mobile – can be used as a stand-alone device or in a smart combination, with rechargeable battery for long
operating time
Defibrillator
Model: Meducore Standard
Manufacturer: Weinmann
Origin: Germany
Specifications:
Device dimensions (W x H x D): 206 mm x 137 mm x 130 mm
Weight, incl. battery 2.2 kg
Operating time Approx. 6 hours monitoring
Power supply 12 to 15 volts
Charging time from 0 to 95% Approx. 3.5 h
Display: 5.7” TFT color display; Resolution:320x240 pixels
Information displayed
ECG monitoring
• Patient connection: 4-pole ECG cable
• Heart rate 30 to 250 /min
• Feed rate 25, 50 mm/s
Saving of operating data
• Internal memory approx. 7 h
• SD card (2 GB) approx. 150h
Defibrillator
• Wave shape biphasic, current-limited, impedance-compensated
• Energy level can be set from 1 J to 200 J
• Charging duration approx. 8s (200 J)
• Shock sequence can be programmed: constant or escalating
• Patient impedance maximum 200 Ω, minimum 5 Ω
AED mode
• Report: ERC 2015
• Duration of analysis 8s
• Voice prompts adjustable: only applies to defibrillation or cardiopulmonary resuscitation, incl. defibrillation
• Metronome, adjustable 100 /min., 110 /min., 120 /min., deactivated
• Duration between cardiac rhythm analyses can be set: 75 to 300 s
• ECG display can be set: Pad lead
• Plethysmogram can be set
Alarm system
• Alarm limits regulated: heart rate (30 to 250 /min), pulse rate (30 to 250 /min), oxygen saturation (85 to 99%)
• Pausing of audio alarm output, adjustable 1, 2, 5,10 minutes or permanently
• Alarm sound cancellation
Start mode Regulated AED mode, monitor mode