Endograph DC is a new, hi-frequency intraoral unit, entirely engineered and manufactured by Villa Sistemi Medicali. It is characterized by an innovative control unit with a large area display which is one of the most distinguishing part of the machine. Endograph DC opens a new era for intraoral units: a new era characterized by peerless convenience, easiness of use and innovative design.
Dental X Ray
Model : Endograph DC
Manufacturer : Villa Sistemi Medical
Origin : Italy
Tubehead Features
High voltage circuit type : High frequency (DC)
High voltage frequency : 30 kHz
Rated output voltage : 60 - 65 - 70 kVp ± 8%
Tubehead current : 6 mA
Maximum deviation of output current : ± 10%
Nominal tubehead power : 420 W
Total filtration : 2.5 mm Al eq. @ 70 kV, HVL > 2 mm Al eq.
Interval between exposures (duty cycle) : 1:60
Minimum focus to skin distance : 20 cm , 30 cm with optional collimator cone
X-ray field (at collimator tip) : Circular, diameter 6 cm / Rectangular, 3 choices (20x30mm, 25x35mm, 35x45mm) with optional diaphragm
Leakage radiation @ 1m : < 0.25 mGy/h
Technical factors for leakage radiation : 70 kV, 6 mA, 1 s, 60 exp/h
Tube insert type : OX 70-5
Focal spot : 0.5 (IEC 336)
Anode material : Tungsten
Anode angle : 19°
Inherent filtration : 0.5 mm Al eq. @ 70 kV
Anode thermal capacity : 7 kJ
Timer Features
Exposure times : From 0.01 to 2s in 36 steps
Exposure times control : Microprocessor controlled exposure times
Timer accuracy : ± 5% or ± 2ms (whatever is greater)
Timer display : LCD graphical display 84x45mm.
Resolution: 240x128 pixels
Exposure button : With extensible cable (3m)
Number of anatomic programs : 882 pre-set times
Choice of patient size : 3 choices (small, medium, large)
Choice of anatomic region
- Incisors (upper and lower)
- Canines (upper and lower)
- Premolars (upper an lower)
- Upper molars
- Lower molars
- Anterior bite-wing
- Posterior bite-wing
Selection type of receptor : 3 choices
- Films (up to 5 different speeds)
- Phosphor plates
- Digital sensors
Dose display : Irradiated dose (mGy) is estimated by software according to the exam exposure factors. Reading accuracy: ±20%
Variation of times for collimator cone = 30cm : A keyboard selection, with display indication, allows to change automatically the exposure times when the
30 cm cone collimator is used
Visualization of number of exposures : Yes
External connections Provisions for connection of:
- Remote x-ray pushbutton
- Ready status signal lamp
- X-ray emission signal lamp