General
12.1” high resolution(800X600) with maximum 10 waveforms
Standard Configuration : ECG, SpO2, NIBP, Resp, 2Temp, 2IBP
6 Kinds of Virtual Screen
3CH ECG : Full 7 ECG & ST Segment analysis
12CH EKG : Full 12 ECG & ST Segment analysis (Optional)
13 Arrhythmia analysis & Pacemaker detection
Saves tabular and graphic trend data for 7 days
Multi high-end Functions
– Drug Dose Calculation, ECG Recall, Mini Trend, ICO, Color change, NIBP STAT
– NIBP VENOUS STAT, OXY-CRG, Patient information, Temp
Easy S/W upgrade with SD Card (Max 8G byte)
Durable, and light Long Li-ion battery operation
– Std 1Pack : 2hrs / 2Pack : 4hrs
LAN/Wireless LAN Connection
Respironics’s EtCO2 Mainstream (C5), Sidestream (LoFlo)
Phaseln’s Multi-gas Mainstream(IRMA AX+), Sidestream(ISA AX+, OR+)
BIS(qCON) Measurement
HRV, APG Detection
Touch Screen(Option)
VGA OUT(Option)
Supreme Standard configuration and extendibility
– Standard : ECG, SpO2 NIBP, Respiration, 2Temp, 2IBP
– Optional : Standard + 4Temp, 4IBP, ICO, EtCO2, 12CH EKG, Touch Screen, 3CH Printer, VGA OUT, ECG OUT
Patient Monitoring
Model: VP-1200
Manufacturer: Votem
Origin: Korea
Specifications:
Display: 12 inch (800 X 600);
Up to 10 waves
7 ECG wave display(I, II, III, aVR, aVL, aVF, V);
Parameters: ECG, SpO2, NIBP, IBP, Resp, Temp;
Memory storage: 7 days, save 20 event data;
Tabular and Graphic Data interval: 1, 5, 15, 30 min và 1 hour.
ECG
Lead: 05 lead;
Gain: 2.5, 5, 10, 20, 30, 40, Auto (mm/mV);
Sweep speed: 6.25, 12.5, 25, 50 mm/sec
HR Range: 0 - 300 bpm (± 2 bpm)
Pacemaker Mode:(PNF, PNC);
Arrhythmia (TAC, BRD, PVC, VTAC, ASY, BGM, TGM, VENT, VFIB, CPT, TPT, MIB, RonT)
ST Analysis: -9.9 mm đến +9.9 mm;
Resolution: 0.1 mm
SpO2
Range: 0 to 100%;
Accuracy: ±2% (70 - 100%); ±3% (50 - 69%); Unspecified (0 - 49%)
Pulse range: 0 - 300 bpm;
Accuracy: ± 2bpm (0 - 240 bpm); ± 3bpm (241 - 300 bpm)
Low perfusition: 0.1 % ;
Time setting: 2s (WOT); ≥ 10s (percentage display SpO2)
Gain: 0.25, 0.5, 1, 2, 3, 4, Auto(mm/mV);
Sweep speed: 6.25, 12.5, 25, 50 mm/sec
Respiration
Range: 0 - 200 rpm (± 2% or ± 2bpm);
Apnea: Off, 10 - 40 sec;
Waveform: 0.25, 0.5, 1, 2, 3, 4, Auto
NIBP
Method: Oscillometric;
Measurement: 0 to 300 mmHg (Adult); 0 to 150 mmHg (Neonate);
NIBP STAT: 5 - 15 min;
NIBP Venous Stat: 50 - 200 mmHg
IBP
Channel: 2 ;
Range: -50 to 350mmHg;
Accuracy: ±1% (mmHg)
TEMP
hannel: 2 ;
Range: 0 to 50.0°C;
Accuracy: ± 0.1°C (25.0 to 50.0°C), ± 0.2°C (0 to 24.9°C)
Interface: RS-232, LAN